Page #31 background imageLoading...
Page #31 background image
Menu mode
Changing transmitter upper parts
73
Remove slave from the system bus –
Display then reads:
or:
Connect new slave – Display:
Change other slaves in the same way.
The displayed number of slaves changed increases after each change –
Display e.g.:
When changing is completed:
Press [ u ] key to quit function and display menu option again.
The following error message may appear when changing transmitter upper
parts:
Only slaves of the same type can be exchanged.
Display if a slave is replaced by a different type (e.g. a CO transmitter by
an NH
3
transmitter or a relay module):
The function cannot be executed in this case.
Slave type are diff. !
0 Slaves changed !
MENU:
CHANGE SLAVES
Connect slave !
0 Slaves changed !
New slave detected !
1 Slaves changed !
New slave detected !
5 Slaves changed !
Connect slave !
Line disconnected !