XS series operating manual 
 
XS user manual v0.55  - 111 - 
 ETHERNET Interface 
 
The ETHERNET  function must be selected in the SETUP menu page 2 on the line  
INTERFACE : ETHERNET  
IMPORTANT  
 
The test units of the XS series could be equipped with an ETHERNET interface. This option makes 
possible the integration of any unit of the series in an automatic measurement or test system in 
manufacturing or incoming inspection department. 
 
This interface is accessible through a RJ45 connector provided on the rear panel.  
 
 
 
It is essential to connect the safety loop when using this option with hipot test for example. 
It’s easy to lunch a test from a place, from where it is not possible to check what the unit does. 
 
Connexion examples 
1)Simple connexion between un PC and a unit with a fixed IP address 
 
In this case, you must use a Crossover Ethernet Cable. 
To clearly identify the unit, You must  select an IP address for the unit and the  PC too. 
 
 
 
: 
In example : 
 
PC : 
-  IP Address  : 192.168.0.1 
-  subnet mask   : 255.255.255.0 
 
XS unit: 
-  IP Address   : 192.168.0.2 
-  subnet mask   : 255.255.255.0