XS series operating manual 
 
XS user manual v0.55  - 22 - 
3  Megohmmeter function (insulation resistance 
measurement) 
 
500V megohmmeter function 
 
(MXS500,DXS50/56,DXS500/506,SXS50/56,SXS500/506) 
Measurement voltage   
 
Voltage adjustable from 10V to 500VDC by 1 volt step. 
Accuracy : ±(1% + 1V). 
 
Positive pole of the HV generator is grounded ( earth)  
Dynamic stability for a mains V = ±10%  : >1%. 
The maximum current in the measurement circuit is : 2 mA ±20%. 
The discharge current is limited by a  2.2 k resistor. 
 
Measurement range  
 
Display : 2000 digits 
The measurement range is defined by the following formula : 
 
(Utest / Umax generator ) x 200 G* 
 
* or 2T according to the option 
 
This gives the following board for the 50,100, 250, 500VDC standard voltages  
 
 
Measurement accuracy  
 
2000 digits numeric display with unity indication (K, M, G,T). 
Accuracy (% of the reading, 1U = 1 digit ) : 
 
2 T option and U<=200VDC 
2 T option and U > 200VDC