XS series operating manual
XS user manual v0.55 - 67 -
IMAX and delta I thresholds adjustment during the ramps
Refer in this chapter to the paragraphs IMIN and IMAX threshold adjustment and Delta I level
adjustment how to set the values.
Fig 8.14
Fig 8.15
Control of the display during the ramps
It is possible to have the display of the voltage and/or the current during the ramps, but select
nodisplay of U or I provide a better linearity of the ramps.
Fig 8.16
Power mode for 60335 and 60204 standards
The European standards EN60335 and 60204 demand that the unit be capable to generate one
current of 100mA AC minimum for tests voltages between 500VAC and 4000V AC.
To be conform set the option 335/204 STANDARD on.