XS series operating manual 
 
XS user manual v0.55  - 74 - 
Resistance threshold selection  
 
The unit includes two comparison thresholds making possible to check if the specimen under test 
is good or bad.  
These thresholds can be a resistance or a voltage (according to the requirements of standard 
EN60204-1). 
The H LIMIT defines the maximum resistance value which is allowed for the specimen under test. 
The L LIMIT defines the minimum resistance value in Ohms or Volts that can be reached by the 
specimen under test.  
A specimen is good (PASS) if the resistance is under H LIMIT and over L LIMIT, otherwise the 
specimen is declared bad (FAIL).  
Those thresholds are adjustable from 0 to 1500m or 0.01V to 12.0V. 
 
  Move the reverse video line in front of the H LIMIT line. 
  Press on the RIGHT arrow or ENTER key. 
  Display shows : 
 
 
Fig 9.5 
 
  Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys. 
  Increase or decrease the value with respectively the UP and the DOWN arrow keys (0 1 2 3 4 5 
6 7 8 9.  0 1 2 ...). 
  Enter the threshold with the ENTER key. Warning, the H LIMIT must always be higher than the 
L LIMIT, otherwise an error message is displayed :  
H LIMIT < L LIMIT 
  To set the L LIMIT value, proceed the same way as for the H LIMIT. Warning, the L LIMIT must 
always be lower than the H LIMIT, otherwise an error message is displayed : 
L LIMIT > H LIMIT 
  If you whish to modify the unity (for instance switch from Ohm to Volt), use the right arrow as 
much  times  as  you  need  to  place  the  unity  on  reverse  video.  Then  change  it  with  UP  and 
DOWN arrows. 
 
Warning : when changing the unity (ohm to volt), you must enter again the two thresholds.