Installation Serinus 51 User manual 1.0
14
2.4 Transporting/Storage
Transporting the Serinus should be done with great care. It is recommended that the
packaging the Serinus was delivered in should be used for any transport as it ensures
minimal damage to the instrument. When transporting or storing the instrument the
following points should be followed:
• Turn off instrument and allow to cool down
• Remove all pneumatic, power and communication connections*
• Remove instrument from rack
• Replace red plugs into pneumatic connections
• Place instrument back in plastic bag with desiccant packs and seal bag (Ideally
the bag it was delivered in)
• Place back in foam and box instrument was delivered in if possible, if not find
some equivalent packaging which provides protection from damage
• Instrument is now ready for long term storage or transportation
Note: After transport or storage, instrument
must be set up and calibrated see section 2.3.4
*if storing over long period (6 months) turn the battery off by switching the following switch on the main
processor board (Figure 6) to the left
Figure 6. Switching battery off