Calibration Serinus 51 User manual 1.0
44
7. Click the “Telnet Configuration” tab and click “Connect” (Figure 14). Leave the
IP address and port as is. Afterwards, press Enter to enter Setup Mode. This will
list the various parameters set up in the Serinus and will give you menu options to
change them.
Figure 14 Device Installer - Telnet Configuration
8. On the “Change Setup” menu, Press 0 and then Enter to Open the Server setup
a) Press Enter to accept current IP Address value or type in new value (must
press enter after each number is entered)
b) Press “Y” to set Gateway address (this value should be obtained from your
network administrator) and enter Gateway address
c) Set the number of Netmask bits e.g.
24 255.0.0.0
16 255.255.0.0
8 255.255.255.0
d) When it asks you to “Set password” select N
9. On the “Change Setup” menu, Press 1 and then “Enter” to open the Channel 1
setup (below).
Change Setup:
0 Server
1 Channel 1
3 E-mail
5 Expert
6 Security
7 Defaults
8 Exit without save
9 Save and exit Your choice ? 1
10. Change the settings to match those in brackets below as they appear. The value
in brackets indicates its current configuration. If the value should be left
unchanged, simply press Enter without any value being entered (below).
Baudrate (38400) ?
I/F Mode (4C) ?
Flow (00) ?
Port No (32785) ?
ConnectMode (C0) ?
Send '+++' in Modem Mode (Y) ?
Auto increment source port (N) ?