EasyManua.ls Logo

Ecotech Serinus 51 - Figure 20 Plugged Sample and Calibration Ports

Ecotech Serinus 51
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus 51 User manual 1.0
Service
57
Figure 25 Plugged sample and calibration ports
3. Enter the main menu status menu pressure menu and compare the
ambient pressure to the inst gas pressure
4. Allow 1-2 minutes for the pressures to stabilize, check that the gas flow
= 0, if this is not the case then there is a leak
5. If the two pressures are within 10% of each other then the leak check has
passed.
6. If the pressures fluctuate with channel switching or are more than 10%
different, perform pressure calibration (section 4.5), after this is
completed repeat quick leak check, if this again fails perform advanced
leak check (section 6.4.6).

Table of Contents

Related product manuals