Calibration Serinus 51 User manual 1.0
38
Figure 8 Excel graph of Multipoint Calibration
12. Accept the calibration if the following are found:
• The gradient (m) falls between 0.98 and 1.02
• The intercept (b) lies between ±0.3.
• The correlation (R
2
) is greater than 0.9995
13. Reject the calibration if the above criteria are not met. If the calibration fails
perform a leak check (section 6.4.5), check zero air scrubbers or check
troubleshoot guide for possible errors (section 7).