Service Serinus 51 User manual 1.0
64
Dessicant, 5 gram pack (2 required) C050014
O-ring, Test Plug, Test lamp O010021
Fitting kynar male branch tee F030033-01
Washer neo, .174x .38x.016 F050041
O-ring BS115, viton O010004
O-ring BS015, viton (3 required) O010023
O-ring, reaction cell cover plate ORI-1019
Tygon tubing (3ft) T010011
Other consumables (not listed in maintenance kit)
Filter paper Teflon 47MM pack of 50 F010006-01
Filter paper Teflon 47MM pack of 100 F010006-01
Charcoal activated, 1KG bottle A-ECO-1035
Orifice, 14 mil H010043-13
Orifice, 20 mil H010043-19
Lamp Assembly, UV C020076
External pump repair kit (suite 607 pump) P031001
Silicone heatsink compound C050013
SO2 Scrubber H040011-01
Molycon Assembly, Serinus, H2S H011105-03
Bypass Orifice W/O-Ring Groove, 12ML H010043-11
6.6 Updating firmware
USB memory stick update
1. Turn instrument off
2. Place USB memory stick with new firmware on the Front panel USB
port
3. Turn instrument on and immediately press the hash key multiple times
until the following screen appears
Vx.x Bootloader
Press ‘1’ to enter Bootloader
4. Select option 3, (Upgrade from USB memory stick) press 3 on keypad
5. Once update is completed.
6. Press 9 to start the analyzer with new firmware