5.5.4 Remote terminal
The remote terminal tab is a diagnostic tool used to check instrument
operation and parameters. The remote terminal is used in a similar way to
“ping” a computer; it is used to ensure communications are working properly.
Firstly the Downloader program must be connected to the instrument and the
green connect button in the top left hand corner must be clicked. The remote
terminal tab consists of 3 different sections:
• Advanced Protocol
Within this section the user selects the parameter number (from within
the advanced protocol) then clicks the get button to receive that
parameters value
• EC9800 Protocol
Within this section a command can be sent to the instrument using the
EC9800 protocol. Simply type in the text command in the field and
click the send button.
• Received Data
This field contains the value or the parameter that has been requested.
Values can be cleared by pressing the clear button
Figure 20 Serinus downloader - Remote Terminal tab