Operation Serinus 51 User manual 1.0
20
3.5.2 Analyzer State
Displays the status of various parameters that effect instrument measurement and
various functions. There are 4 sub menus within this menu and 3 non-editable fields
described below.
Event Log
This field enters a screen with a log of all the events that the instrument performs.
These events include calibrations, errors, backgrounds, warnings. This log is stored on
the removable USB flash memory.
Firmware Ver.
This field displays the firmware version currently in use on this analyzer. This can be
important when performing diagnostics and reporting back to the manufacturer.
Power Failure
This field displays the time and date of the last power failure occurred (or power was
disconnected from analyzer)
3.5.2.1 Status
The Component status menu contains a list of the main components
within the instrument and their status.
Cell Temp
The cell temperature must be within ± 10% of the heater set point
(section 3.5.3)
Cooler Status
Status of the PMT cooler,
Sample Gas Flow
Indicates whether the instrument has flow or not.
A/D Input
A reference voltage is sent to analog to digital input chip, this field will
display a pass or fail indicating if the board is working or not.
Chassis temp
Displays whether the Chassis temperature is within the acceptable limits
Reference Voltage
Checks that the background is performing within acceptable limits
Pass/Fail.
Lamp/Source
Checks if the Lamp current is between 20-50 mA. Pass if within these
limits, Fail if outside.
USB Key
Detects whether a USB key is plugged into the USB port and instrument
is logging