Serinus 51 User manual 1.0
Operation
21
Flow Block Temp (Optional internal pump only).
If you have the optional internal pump this temperature must be within
10% of the heater set point (Section 3.5.3) to keep a constant accurate
flow.
3.5.2.2 Temperatures
Set point
(CELL)
Sets the temperature that the heated components will be regulated at
including the cell.
Set point
(CONVERTER)
Sets the temperature that the converter will be regulated at.
Temperature Units
Editable field to allow the user to change the current temperature units of
the analyzer.
Cell
Temperature of the reaction cell.
Converter
Temperature of the thermal converter.
Flow block (optional internal pump only)
If an internal pump is installed, this field displays the current
temperature of the flow block.
Chassis
Displays the temperature of air inside the chassis, measured on the
microprocessor board.
PMT Cooler
Temperature of the cooled PMT block
3.5.2.3 Pressures & Flow
Pressure units
Select the units that pressure will be displayed in (torr, psi, bar, atm, kPa)
Ambient
Current ambient pressure (outside the analyzer).
Cell
Current pressure within the measurement cell