EasyManua.ls Logo

Ecotech Serinus 51 - 6.3 Maintenance schedule

Ecotech Serinus 51
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Serinus 51 User manual 1.0
54
6.3 Maintenance schedule
Table 1 Maintenance Schedule
Interval *
Task performed
Page
Weekly
Check inlet particulate filter, replace if full/dirty
55
Check sample inlet system for moisture or foreign
materials. Clean if necessary
Perform precision check
40
Monthly
Check fan filter, clean if necessary
55
Perform span calibration
35
Check date and time is correct
23
6 Monthly
Check the zero air scrubber
58
PMT Desiccant packs
59
Perform multi-point calibration check
36
Yearly
Replace DFU filter
56
Replace sintered filter and orifice (only if necessary)
Check UV lamp alignment (or as required)
60
Perform pressure check
61
* Suggested intervals for maintenance procedure may vary with sampling intensity and environmental
conditions.

Table of Contents

Related product manuals