Serinus 51 User manual 1.0
Operation
15
3
Operation
3.1 Warmup
As the instrument is turned on it will adjust and calibrate itself ready for monitoring.
No measurements are taken during the warm-up.
When the main screen appears, the instrument will display the following actions being
taken (bottom of the screen):
High voltage check:
Instrument setting upper voltage limit for measurement
Lamp Stabilize
Adjusts the lamps current (27mA) for a stable (reference voltage) signal/output (10-15
minutes).
Ref stabilize
Adjusts the detector to set its maximum voltage output and create a stable output
signal.
After this warm-up has completed the instrument will immediately begin taking
measurements as seen in section 3.4.
3.2 General operational information
The Serinus is operated with the use of 3 sets of buttons. There are the selection
buttons, scrolling buttons and keypad.
Figure 7 Serinus front panel
(1) Selection buttons
The selection buttons will perform the function specified directly above it on
the screen i.e. quick menu, main menu, open, back, select, start etc