Serinus 51 User manual 1.0
 
Operation
 
27 
Save parameter List 
Name the file which the parameter settings will be downloaded into (at 
that time), used for diagnostics  
Load Configuration 
Loads instrument EEPROM setting from USB key (including calibration 
and communication settings, units, instrument gain etc). 
Auto-Backup Interv 
The Param list and config file are saved onto USB for the duration set in 
this field 
Load Auto-Backup Config 
Loads up the last auto backup configuration (best used when 
configuration has been changed in error) 
 
----------------------------------------------------------------------------------- 
Instrument 
This field allows the instrument to be set to either “Online” (Normal 
instrument operation) or “Offline” (data is not valid, as service work etc 
is being performed). This field is used to change the instrument into 
“Offline” when service work is being performed. 
Next service due 
An editable field where the user can set the date when the next 
instrument service is required. Section 6.3 has a recommended 
maintenance schedule that may be used as a basis for the interval entered 
above. This value is also displayed as a non editable field in the quick 
menu section 3.5.1.  
Rebuild event Log Index 
This feature will rebuild the logs on the USB key. This is necessary 
when any changes to the files or folders on the USB key have been made 
(including within a file). Rebuilding the log index will re-index all the 
contents and thus allow the instrument to view error logs. 
Safely remove USB stick 
This feature must be activated to safely remove the USB stick (also 
occurs in the quick menu 3.5.1). 
System Restart 
Activating this will reboot the microprocessor. This action is useful 
when uploading new firmware or if tests are being carried out. 
----------------------------------------------------------------------------------- 
Avg conc.  
Displays the SO
2
 / H2S average concentration