Operation Serinus 51 User manual 1.0
30
Digital Inputs
This menu item allows the user to set which pin the power source is
located on and also displays the pins that digital inputs are located on.
Note: Changes performed with this
item will disrupt the normal logging
Digital Outputs
This menu item allows the user to view the pins that digital outputs are
located on
Note: Changes performed with this
item will disrupt the normal logging
Analog Inputs & Outputs
Displays the pins that analog inputs are located on and allows the user to
set the analog outputs
Note: Changes performed with this
item will disrupt the normal logging
3.5.6.2 Calculation factors
The Calculation factors screen provides a list of calibration values used
by the instrument to generate valid data. The operator can collect these
values and report them on calibration reports as part of their quality
assurance.
Instrument gain
A multiplication factor used to adjust the concentration measurement to
the appropriate level (set at calibration)
Zero Offset
This field displays the offset created from a zero calibration, this is the
concentration measured from zero air and is subtracted from all readings.
Background
The correction factor calculated from the background cycle (used to
eliminate background interferences).
P/T/F Correction
Displays the correction factor applied to the concentration measurement.
This factor corrects for changes in pressure, temperature and flow.
H2S efficiency
The H2S efficiency value is derived from the losses associated in the
selective SO2 scrubber and H2S to SO2 conversion.