Operation Serinus 51 User manual 1.0
32
Data Log setup –GUI below
Data Log setup –GUI
This field works in tandem with the “Data Log setup –Quick” field and
is used when the parameter number is not known. The user can enter
each logging parameter field (P1-P15) and scroll through all the
parameters that can be logged. Once the parameters to be logged has
been filled out, the parameter numbers can be copied from within the
“Data Log Setup –Quick” field for future use.
Data Log Interval
Set the interval (length) of time that measurement data will be logged
over i.e. 1 minute data, 5 minute data etc
3.5.7.3 Serial Communications
Multidrop ID
The ID address of the analyzer when Multidrop RS232 communications
is used.
Baud
(RS232 #1)
Set the Baud rate that the instrument will communicate over
(1200, 2400, 4800, 9600, 14400, 19200, 38400)
Protocol
(RS232 #1)
Set the protocol used for communication either EC9800, Bavarian or
Advanced
Baud
(RS232 #2)
Set the Baud rate that the instrument will communicate over
(1200, 2400, 4800, 9600, 14400, 19200, 38400)
Protocol
(RS232 #2)
Set the protocol used for communication either EC9800, Bavarian or
Advanced
Delay
(RS232#2)
When set to “enabled” will delay the serial communication through the
RS2323 #2 port by approximately 0.25 seconds, this is used in systems
that cannot cope with the rapid response of the Serinus instruments.
When set to “disabled” communication will proceed as normal.
3.5.7.4 Digital Inputs Menu
Assign the digital input 0-7