Serinus 51 User manual 1.0
Appendix
77
84 Reference Voltage
85 Calibration State 0 = MEASURE
1 = CYCLE
2 = ZERO
3 = SPAN
86 Primary Raw Concentration (before NOx background and gain)
87 Secondary Raw Concentration (before NOx background and gain)
88 NOx Background Concentration (before gain)
89 Calibration Pressure
90 Converter Efficiency
91 Multidrop Baud Rate
92 Analog Range Gas 1
93 Analog Range Gas 2
94 Analog Range Gas 3
95 Output Type Gas 1 1=Voltage
0=Current
96 Output Type Gas 2 1=Voltage
0=Current
97 Output Type Gas 3 1=Voltage
0=Current
98 Voltage Offset /Current Range Gas1 0=0% or 0-20mA
1=5% or 2-20mA
2=10% or 4-20mA
99 Voltage Offset /Current Range Gas2 0=0% or 0-20mA
1=5% or 2-20mA
2=10% or 4-20mA
100 Voltage Offset /Current Range Gas3 0=0% or 0-20mA
1=5% or 2-20mA
2=10% or 4-20mA
101 Full Scale Gas 1 5.0 Volt Calibration value for Analog Output 1
102 Full Scale Gas 2 5.0 Volt Calibration value for Analog Output 2
103 Full Scale Gas 3 5.0 Volt Calibration value for Analog Output 3
104 Zero Adjust Gas 1 0.5 Volt Calibration value for Analog Output 1
105 Zero Adjust Gas 2 0.5 Volt Calibration value for Analog Output 2
106 Zero Adjust Gas 3 0.5 Volt Calibration value for Analog Output 3
107 Negative 10V Supply
108 NA
109 NA
110 Instrument State
111 CO Linearisation Factor A
112 CO Linearisation Factor B
113 CO Linearisation Factor C
114 CO Linearisation Factor D
115 CO Linearisation Factor E
116 Instrument Units 0= PPM
1=PPB
2=PPT
3=mG/M³
4=µG/M³