sunderwoodAug 25, 2025
What to do if EDAN iM80 displays 'Art ZERO FAIL'?
- SSabrina LeeAug 26, 2025
If the EDAN Monitor displays 'Art ZERO FAIL', make sure that the transducer is not attached to the patient.
What to do if EDAN iM80 displays 'Art ZERO FAIL'?
If the EDAN Monitor displays 'Art ZERO FAIL', make sure that the transducer is not attached to the patient.
What to do if EDAN Monitor displays 'Art SENSOR OFF, FAIL'?
If the EDAN Monitor displays 'Art SENSOR OFF, FAIL', ensure the transducer is not off, and then proceed with zeroing.
What to do if EDAN iM80 Monitor displays 'IN DEMO, FAIL'?
If the EDAN Monitor is displaying 'IN DEMO, FAIL,' make sure that the monitor is not in DEMO mode. Contact a service technician if necessary.
How to fix EDAN Medical Equipment when LCD has no display, power indicator is off, and the fan doesn’t run?
If your EDAN Medical Equipment's LCD screen is blank, the power indicator is off, and the fan isn't running, start by replacing the fuses. If that doesn't solve it, the power board might be defective and need replacement. Also, check for any short circuits in other parts and fix them.
What to do if my EDAN Medical Equipment is abruptly switching off?
If your EDAN Medical Equipment is abruptly switching off, check the power supply and earth system, especially if there was a strong high voltage event like a lightning strike. Also, a bad power supply can cause this, so consider replacing it. A defective main control board or bad connection could also be the issue; in these cases, replace the main control board or repair/replace the connectors.
Why does my EDAN Medical Equipment LCD have no display or is black when working?
If the LCD on your EDAN Medical Equipment shows no display or is black while working, the issue might be with the LCD backlight board, which would need replacing. Alternatively, the LCD cable could be defective, requiring repair or replacement. The main board or power board could also be defective, requiring replacement.
How to troubleshoot network connection issues with EDAN Medical Equipment?
If your EDAN Medical Equipment can't connect to a network, check and repair the network wire and HUB box. It could also be due to an overlapped device number in the network, so try changing the device number of the monitor. In some cases, the main control board may be defective and need replacement.
What to do if EDAN iM80 shows 'PRESSURE OVER RANGE, FAIL'?
If the EDAN Monitor displays 'PRESSURE OVER RANGE, FAIL,' ensure the stopcock is vented to the atmosphere. If the problem continues, please contact a service technician.
Why is the touch screen not working on my EDAN Medical Equipment?
If the touch screen on your EDAN Medical Equipment isn't working, first check if it's disabled in the system setup and enable it if necessary. Then, ensure that the cables between the keyboard and touch screen are securely connected. If these steps don't resolve the issue, the touch screen may be damaged and require replacement.
Why is there no audible alarm on my EDAN Medical Equipment?
If no audible alarm is activated on your EDAN Medical Equipment, first ensure that the audible alarm is not temporarily disabled and switch it on if necessary. If the alarm is enabled but still not working, the loudspeaker or the wire may be defective, requiring replacement.