E-connect S

Questions and Answers

  • P
    Pamela BensonSep 9, 2025
    What to do if motor does not run when file is inserted in canal in Eighteeth Dental equipment?
    • S
      sarahkelleySep 9, 2025
      If your Eighteeth Dental equipment motor does not run when the file is inserted in the canal, check that the measure wire and lip hook are properly connected. Also, ensure that the auto-start function is turned on, if needed.
  • C
    Carol BrownSep 8, 2025
    Why is the power LED on my Eighteeth E-connect S Dental equipment charge base not lighting?
    • J
      Jamie SmithSep 8, 2025
      If the power LED on the Eighteeth Dental equipment charge base isn't lighting, make sure you're using the original adapter and that it's properly connected to both the base and a working outlet. Verify there is electricity in the outlet.
  • V
    Vanessa SheltonSep 7, 2025
    What to do if the motor doesn’t rotate on my Eighteeth E-connect S Dental equipment?
    • B
      Brett FitzpatrickSep 7, 2025
      If the Eighteeth Dental equipment's motor doesn't rotate, first, ensure you're not in M0 mode, which is for stand-alone apex locator function, and switch to M1-M10. Next, check if the contra-angle is clogged and clean or replace it. Finally, examine for any error warnings indicating motor protection or breakage.
  • K
    Karen HamiltonAug 22, 2025
    What to do if motor doesn't run when file is inserted in the canal in Eighteeth E-connect S Dental equipment?
    • J
      Joshua ElliottAug 22, 2025
      If the motor in your Eighteeth dental equipment doesn't run when the file is inserted in the canal, consider these potential causes: * The measure wire might not be properly connected, so check the connection. * The lip hook may not be correctly hooked in the corner of the patient’s mouth; verify its connection. * The auto-start function could be off. If necessary, turn the auto-start function on.
  • M
    Martin WilsonSep 4, 2025
    How to turn on sound in Eighteeth E-connect S Dental equipment?
    • R
      rodney81Sep 5, 2025
      If there is no sound coming from your Eighteeth dental equipment, the beep volume might be set to 0. Set the beep volume to 1, 2, or 3.
  • D
    David GonzalezSep 3, 2025
    Why motor alternates between forward and reverse rotation in Eighteeth E-connect S?
    • R
      romerolisaSep 3, 2025
      If the motor alternates between forward and reverse rotation in your Eighteeth dental equipment, the operation mode setting might be set to REC or ATC. Change the setting if it’s not expected.
  • L
    lisa93Sep 2, 2025
    Why motor speed changes spontaneously in Eighteeth E-connect S?
    • M
      mitchellryanSep 2, 2025
      If the motor speed changes spontaneously in your Eighteeth dental equipment, the apical action setting might be set to Slow Down. Change the setting if it’s not expected.
  • M
    Mary RobertsSep 1, 2025
    Why canal measurement is unstable in Eighteeth E-connect S?
    • K
      Katherine HartSep 1, 2025
      If the canal measurement is unstable with your Eighteeth dental equipment, it might be due to a complex root canal environment. Check the situation of the root canals.
  • J
    Jordan CharlesAug 31, 2025
    Why the motor does not reverse in Eighteeth Dental equipment?
    • J
      Jason DavisAug 31, 2025
      If the motor in your Eighteeth dental equipment isn’t reversing, here are some potential reasons: * The R.L mode might be set. Change the setting if it’s not expected. * The torque reverse setting might be too high. Change the setting if it’s not expected. * The apical action setting might be set to Stop or OFF. Change the setting if it’s not expected.
  • D
    Deborah MontgomeryAug 29, 2025
    Why I cannot make a measurement with Eighteeth E-connect S?
    • M
      Madison MarksAug 30, 2025
      If you're unable to make a measurement with your Eighteeth dental equipment, check the following: * The measure wire, lip hook, or file clip might not be properly connected. Check the connection. * There might be a lack of electrical conductivity between the shank and the file. Use a file that has conductivity. * The root canals might be unsuitable. Check the root canal environment.

Summary

3. Before Use

4. Installing the E-connect S

4.1 Installation of the Contra Angle

Instructions for securely attaching the contra-angle to the handpiece, including rotation options.

5. Use Interface

6. Setting

6.2 Setting Parameters

Instructions for adjusting speed, torque, and other operational parameters for specific treatments.

7. Operation

7.2 Motor Operation

Details on operating the motor for dental procedures, including speed and torque indications.

7.3 Apex Operation and Not Suitable Condition

Covers apex locator usage and identifies root canal conditions that may affect measurement accuracy.

8. Maintenance

Autoclavable Components

Lists the specific parts of the E-connect S that can be safely autoclaved for sterilization.

Autoclavable Procedure

Step-by-step instructions for cleaning, disinfecting, lubricating, and sterilizing components.

9. Error Warning

Overload and Overheat Warnings

Explains overload and overheat conditions and the necessary actions to take.

Hardware and Motor Faults

Describes fault conditions related to the handpiece hardware or motor, requiring distributor contact.

10. Troubleshooting

Power and Charging Issues

Addresses problems related to the device not turning on or the charge LED not lighting up.

Motor and Apex Function Issues

Resolves issues like the motor not rotating, not stopping, or not reversing correctly.

Measurement and Sound Issues

Covers problems with unstable measurements, no sound, or alarms during operation.

Eighteeth E-connect S Specifications

General IconGeneral
CategoryDental equipment
Adaptive Torque ReverseYes
Apex Locator IntegrationYes
DisplayOLED
BatteryRechargeable Lithium-ion
Weight150 g
TypeEndodontic motor
Torque Range0.4 Ncm - 5.0 Ncm
Speed Range100 rpm - 1200 rpm
Battery Capacity2000 mAh