Technical specifications
11.7 Climatic conditions
SCD 19102 / SMD 19102
Instruction Manual, 10/2011, A5E01588521A-004
93
11.7 Climatic conditions
Operation
Temperature range 5 °C ... 40 °C ambient temperature
Temperature gradient Max. 5 °C/h, without condensation
Air pressure 1053 ... 540 hPa
Transport and storage (packed)
Temperature range -20 °C ... +60 °C ambient temperature
Temperature gradient Max. 5 °C/h, without condensation
Air pressure 1053 ... 540 hPa
11.8 Mechanical requirements
Operation
Vibration To EN 60068-2-6
10 ... 58 Hz at ± 0.075 mm deflection
58 ... 500 Hz at 10 m/s²
Shock To EN 60068-2-27 (single impact)
150 m/s², 6 ms
Under operating conditions, the device must not
be exposed to continuous or repeated shocks.
Packed unit
Vibration To EN 60068-2-6
5 ... 9 Hz at ± 3.5 mm deflection
9 ... 500 Hz at 10 m/s²
Shock To EN 60068-2-27 (single impact)
250 m/s², 6 ms (in storage packaging)
to EN 60068-2-29 (continuous shocks)