Technical specifications
11.9 Safety specifications
SCD 19102 / SMD 19102
94 Instruction Manual, 10/2011, A5E01588521A-004
11.9 Safety specifications
Safety standards
IEC/EN 60601-1 (Second Edition)
IEC/EN 60601-1 (Third Edition)
CAN/CSA C22.2 No. 601.1-M90
CAN/C22.2 No. 60601-1-08
Protection class Protection class I
Degree of protection
according to DIN 40050
IP20
Conformity CE to MDD 93/42/EEC (Class I)
CAUTION
Subsequent mounting of a stand
The SCD / SMD 19102 C and SCD / SMD 19102 CP displays are delivered without a
stand. To ensure that the complete system (display and stand) complies with the EN 60601
standard, a subsequently mounted stand must also comply with the EN 60601 standard.
11.10 Electromagnetic compatibility
Interference voltage/interference noise EN 60601-1-2
Voltage fluctuations EN 610004-11
Burst on power cables EN 61000-4-4
1 kV
Surge on power cables EN 61000-4-5
1 kV symmetrical,
2 kV unsymmetrical
Static discharge on casing parts (ESD) EN 61000-4-2
8 kV air, 4 kV contact
RFI EN 61000-4-3
80 MHz ... 2.5 GHz,
3 V/m 80 % AM 1 kHz
Noise immunity EN 61000-4-6
150 kHz ... 80 MHz
3 V/m 80 % AM 1 kHz
Magnetic constant fields EN 61000-4-8
Max. 4000 A/m
Magnetic alternating fields EN 61000-4-8
Max. 10 A/m