EasyManua.ls Logo

Eliwell EWCM 4150 - Page 22

Eliwell EWCM 4150
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
* If inputs AI3 and AI4 are not set as DI, parameters CF25 and CF26 parameters must be set to 0. Failure to observe
this rule may result in malfunctions.
** If inputs AI3 and AI4 are set as DI, parameters CF14 and CF15 parameters must be set to 0.
*** The unit of measurement (U.M.) is selected based on parameters CF02 and CF03 and parameters UI22 (C°/F°) and
UI23 (Bar/Psi).
Inputs AI3 and AI4 are configurable as indicated in the following table (CF02 ... CF03):
Value Type Description
0 None Probe not configured
1 DI Probe configured as potential-free contact digital input
2 NTC NTC probe range -50.0°C ÷ 99.9 °C
3 4-20mA Analogue input 4-20 mA
4 0-10V Analogue input 0-10 V
5 0-5V Analogue input 0-5 V
Notes:
If an input is configured as NTC, the corresponding parameters are always displayed with the “thermometer” icon.
(UI22=0/1; U.M.= C°/F°)
If an input is configured as 4-20mA, 0-10V or 0-5V, the corresponding parameters are displayed with U.M. =Bar if
UI23=0 with U.M.=Psi if UI23=1.
Parameters CF04 ÷ CF07
Indicate the reading scale analogue limit values for inputs configured as 4-20mA, 0-10V, 0-5V. (Inputs 3 and 4 only)
If the input is not configured as input 4-20mA, 0-10V, 0-5V, the end of scale parameters lose meaning.

Related product manuals