The value shows the chosen heating rate in °C/min. (The preheating function is switched off by
entering 0 in this place. Then, the heating occurs with maximum speed after start of the UNI-
software.)
3. Save your entry with Enter.
4. The preheating starts automatically with the next start of the software, if the furnace
temperature is lower than 100°C and the analyzer switch is at position 2.
NOTICE
When the preheating function is activated (by any other value except 0) the software
automatically creates a shortcut at the startup menu of Windows. This enables the start of the
UNI-software with every start of the PC. The shortcut is deleted automatically when the
preheating is deactivated.
5.10.2 BIOS-configuration
In order to start the preheating before the work begins, the BIOS has to be configured to start
the PC automatically at the selected time. (The preheating starts automatically by starting the
UNI-software.)
1. Switch on your PC and enter the BIOS by pressing the corresponding key (e.g. F2).
The normal booting process is interrupted and the BIOS menu appears.
2. Choose Power Management Setup
3. Activate "Resume by Alarm" (enabled).
4. Enter the required days and times when the heating up of the analyzer should be started.
5. Save your entries at Save/Exit with Yes.
6. The booting process will be sustained.