Page #48 background image
48
7.5 O-rings replacement
CAUTION
C10.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
Parts of the analyzer can be very hot.
Use heat protecting gloves.
CAUTION
V0062
Injuries in the form of cuts and other personal injuries
Danger from glass splitters
Injuries in the form of cuts can be caused by damaged sample flasks and
glass splitters.
Replace damaged sample flasks
Do not touch glass splitters with your hands.
The O-rings are only replaced when they no longer adequately seal, due to severe damage or
aging. When removing the old O-rings, be sure that the sealing area of the fittings is not
damaged. Grooves where the old O-rings are installed must be cleaned to be completely free of
grease.
New O-rings should under no circumstances be greased before installing but only after
installation. Otherwise, the O-rings would turn with the glass tubes when trying to remove them
at a later stage.
It is recommended not to grease the upper O-ring of the moisture trap.