EasyManuals Logo

ELTRA CS-580A User Manual

ELTRA CS-580A
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
39
6.3.4 Limestone
NOTICE
Calibrate analyser with calcium carbonate of 12% C.
Temperature
1250 °C or higher.
Weight
300mg
6.3.5 Oil, Asphalt and Rubber
Temperature
1450 °C
Weight
80mg-100mg (fuel oil and other thick oils).
Up to 50mg (thin oil analysis)
6.3.6 Wood
Temperature
1300 °C
Weight:
350mg
Set the minimum analysis time to 50s. Set the comparator level to 20mV.