7.9 Heating elements replacing
DANGER
D11.0005
Mortal danger from electric shock while replacing/installing the heating
elements
Exposed power contacts - High Voltage
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
Set the mains switch of the analyzer to position 0 and pull the
mains plug of the mains socket.
The task of inserting/replacing the heating elements may only be
performed by a trained electrician or Eltra service personnel
A
CAUTION
C13.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
Parts of the analyzer can be very hot.
Wait before start working until the analyzer is cooled down.