21
5.3.2 Autoloader Setup
If the Helios analyser was delivered out an autoloader system, it is configured to be used in automatic mode.
In this case, please install the front cover plate (article number 35400-2001).
5.4 Mains power Connections
DANGER
D3.0005
Mortal danger from electric shock
Exposed power contacts - High Voltage
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
Set the mains switch of the analyzer to position 0 and pull the
mains plug of the mains socket.
DANGER
D4.0005
Mortal danger from electric shock
Damaged power cable
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
Never use a damaged power cable to connect the device to the main.
Check the power cable and the plug for any damage before use.
NOTICE
Installation of the machine
It must be possible to disconnet the machine from the mains at any time.
Install the machine such that the connection for the mains cable is
easily accessible.