EM Test CWS500 D
User manual V 1.07 29 / 34
7. Technical Data
Test Level:
Output power 100W (nominal)
Output impedance 50 Ohm, VSWR 2.0 : 1 nominal
Level approx. -13dBm ... 50dBm
Amplifier Class A, Gain > 52dB
Current 4MHz ... 400MHz max. 700mA ( 100W amplifier, without attenuator, F-130A clamp )
Test frequencies ( synthesizer )
Sinusoidal (CW) 10kHz - 1000MHz
Frequency bands 10kHz - 1000MHz step: 1kHz ( ICD and Quickstart online)
Quickstart setting 10kHz - 9,999MHz step: 1kHz
10MHz - 99,99MHz step: 10kHz
100MHz - 1000MHz step: 100kHz
Modulation
AM Frequency : 1Hz ... 3000Hz
Index : 0%...95%
PM Frequency : 1Hz ... 3000Hz
Duty cycle : 10%...80%
Attenuator
Range 0 dB - -63.5dB
Step level 0.5dB
Timing:
Dwell time general td = 0,3s - 9999.9s
pulse modulation td = 3s - 9999.9s
Rest time tr = 0 / 0,3s - 9999.9s
Output
Direct N type connector
EUT control
BNC input FAIL 1 Fail 1; test will be stopped immediately
BNC input FAIL 2 Fail 2; failure is detected, test continues,
Routines (integrated)
Service Service, Setup
Test routines All test only in remote control with ICD software version >3.00

Related product manuals