21
“Calibration Temp” menu.
To calibrate probe’s temperature enter into “Calibration Temp” menu. A professioanl thermometer is required to obtain a
reliable calibration. From “Menu Calibration” choose “Calibration Temp”.
Note: This procedure assumes that instrument is correctly installed and congured, connected to a working
probe. Calibrate using plant’s temperature otherwise unattended results might occur.
Using an external thermometer read actual temperature and edit related eld “Calib. at”. Conrm by pressing wheel.
To end procedure move cursor on “EXIT” and press wheel to proceed to “Save” request screen. Move wheel on “YES”
to save or “NO” to discard changes.