Option Description
2-Point Calibration Appropriate for gases with two main constituents. Requires two calibration
gases.
3-Point Calibration Appropriate for gases with three main constituents. Use only if recommended by
customer support. For more information, see Three-point calibration.
5. Select Next.
6. Select the number of the calibration that you want to perform (CAL 1, CAL 2, CAL 3, or CAL 4).
The device can store up to four independent calibration ranges. All calibrations must have the same
calibration type (2-point or 3-point). Each calibration can use different calibration gases.
7. Enter data for each calibration gas.
a) Select the calibration gas from the list for Calibration Gas Low and Calibration Gas High. If it is
not listed, select Other.
b) Enter the specific gravity, molecular weight, or relative density of the calibration gas for
Calibration Gas Low and Calibration Gas High.
Purging and calibration Configuration and Use Manual
March 2021 MMI-20020954
38 Micro Motion Gas Specific Gravity Meters (SGM)