3. Select DP Units from the drop-down menu. Select Next.
4. Enter the DP Value corresponding to the desired flow rate simulation. Select Next.
5. Repeat steps Step 1–Step 3 for static pressure (Simulate AP/GP) and process
temperature (Simulate PT), if applicable.
6. Select View Results. Select Next. The simulated flow rate and corresponding flow
properties will be shown. Select Next.
7. Select Exit. Select Next. Leaving the Enable Test Calculation window automatically
returns all process variables fixed by the test calculation method to live process
variable measurements.
4.4.2 Configuring fixed process variables
Mass and energy flow Fast Keys 1, 2, 4
Direct process variable output Fast Keys 1, 2, 3
Under the Fixed Process Variables heading, the user may temporarily set the differential
pressure, static pressure or process temperature to a user defined fixed value for testing
purposes. Once the user leaves the Configure Fixed Variable method, the fixed process
variable will be automatically returned to a live process variable measurement.
4.4.3
Analog output loop test
Mass and energy flow Fast Keys 1, 2, 2
Direct process variable output Fast Keys 1, 2, 2
Under the Analog Output Verify heading, a Loop Test can be performed by selecting the
Loop Test
button. The loop test command verifies the output of the transmitter, the
integrity of the loop, and the operations of any recorders or similar devices installed in the
loop.
4.5 Process variables
4.5.1 Process variable tabs
Mass and energy flow Fast Keys 1, 1
Direct process variable output Fast Keys 1, 1
The Process Variables screen shows a graphical representation of the respective variable.
An example of the Primary Variable tab (as shown in Figure 4-7). The chart on these
Process Variables tabs will begin plotting when the user first navigates to the screen, and
will only continue plotting while the user is viewing this tab. The user may view a larger
version of the chart by selecting the Large Chart button.
Each of the four digital output variables has a screen similar to the one shown in Figure
4-7.
Operation and Maintenance Reference Manual
October 2018 00809-0100-4803
118 Rosemount 3051S Multivariable Transmitter