3 Installation
3.1 Overview
This section contains information that covers installation considerations for Rosemount
3051S MultiVariable
Transmitter (Rosemount 3051SMV). The Rosemount
3051SMV Quick Start Guide is shipped with every transmitter to describe basic
installation, wiring, configuration, and startup procedures. Dimensional drawings for each
Rosemount 3051SMV type and mounting configuration are included in .
3.2 Safety messages
Procedures and instructions in this section may require special precautions to ensure the
safety of personnel performing the operation.
WARNING
Explosions
Explosions could result in death or serious injury.
Installation of device in an explosive environment must be in accordance with
appropriate local, national, and international standards, codes, and practices.
Please review the Product certifications section in the Rosemount
3051 Product Data
Sheet for any restrictions associated with a safe installation.
Before connecting a handheld communicator in an explosive atmosphere, make sure
the instruments in the loop are installed in accordance with intrinsically safe or non-
incendive field wiring practices.
In an explosion-proof/flameproof installation, do not remove the transmitter covers
when power is applied to the unit.
WARNING
Process leaks
Process leaks may cause harm or result in death.
Install and tighten process connectors before applying pressure.
Install and tighten all four flange bolts before applying pressure.
WARNING
Electrical shock
Electrical shock can result in death or serious injury.
Avoid contact with the leads and terminals. High voltage that may be present on leads can
cause electrical shock.
Reference Manual Installation
00809-0100-4803 October 2018
Emerson.com/Rosemount 71

Other manuals for Emerson Rosemount 3051S MultiVariable

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 3051S MultiVariable and is the answer not in the manual?

Emerson Rosemount 3051S MultiVariable Specifications

General IconGeneral
Accuracy±0.04% of span
Pressure RangeUp to 4000 psi (276 bar)
Communication ProtocolsHART, FOUNDATION Fieldbus, PROFIBUS PA
Ambient Temperature Limit-40 to 85 °C (-40 to 185 °F)
Power Supply (Fieldbus)9-32 VDC
Measurement TypeTemperature, Flow
Output4-20 mA HART, FOUNDATION Fieldbus
Temperature Range-40 to 85°C (-40 to 185°F)
Material316L SST, Hastelloy, Monel
ServiceLiquid, gas, steam
Warranty5 years
Measurement Range0-4000 psi
Process Temperature Limit-40 to 120°C (-40 to 248°F)
DiagnosticsAdvanced diagnostics
Certifications/ApprovalsFM, CSA, ATEX, IECEx, SIL 2/3
Process Connection1/2-14 NPT
Electrical Connection1/2-14 NPT, M20
Enclosure RatingIP66, NEMA 4X
Stability±0.1% of URL per 5 years