6.6 Inspection
6.6.1 Visual inspection
Not required
6.6.2 Special tools
Not required
6.6.3 Product repair
The Rosemount 3051SMV is repairable with limited replacement options.
All failures detected by the transmitter diagnostics or by the proof-test must be reported.
Feedback can be submitted electronically at Emerson.com/Rosemount/Report-A-Failure.
All product repair and part replacement should be performed by qualified personnel.
6.6.4
Rosemount 3051SMV SIS reference
The Rosemount 3051SMV must be operated in accordance to the functional and
performance specifications provided in Specifications and Reference Data.
6.6.5
Failure rate data
The FMEDA report includes failure rates and the assumptions for how these failure rates
were derived.
The report is available under Certificates and Approvals at Emerson.com/Automation-
Solutions/Pressure/Rosemount-3051SMV.
6.6.6
Failure values
Safety Deviation (defines what is dangerous in a FMEDA): ±2.0% of analog output span
Transmitter response time: provided in Dynamic performance ambient temperature
effect. Self-diagnostics Test Interval: At least once every 60 minutes
6.6.7
Product life
50 years - based on worst case component wear-out mechanisms - not based on wear-out
of process wetted materials
Reference Manual Safety Instrumented Systems Requirements
00809-0100-4803 October 2018
Emerson.com/Rosemount 147