Table 2-11: Common Custom Units - Totalizer (continued)
Custom unit Base unit Base per custom
Thousands of Standard Cubic Meters (KSCM) StdCum 1000
If conversion factor tables or internet search engines are used to determine the Base per
Custom value, it is important to enter the Custom Unit in the “From” field and the Base
Unit in the “To” Field.
To calculate the Base per Custom value for a custom unit not shown in Table 2-9, see the
example:
Totalizer conversion example
To find the Base per Custom relationship for a custom unit of millions of standard cubic
feet (mmscf) and a base unit of standard cubic feet (StdCuft), input the following:
Custom Unit = mmscf
Base Unit = StdCuft
Because:
1 mmscf (Millions of Standard Cubic Feet) = 1000000 StdCuft (Standard Cubic Feet)
Therefore:
Base per Custom =
= = 1000000
Figure 2-35: Totalizer Custom Units - Totalizer Example
2.7.4
Differential pressure
Mass and energy flow Fast Keys 1, 4, 1, 4
Direct process variable output Fast Keys 1, 4, 1, 1