Base per Custom = = = 1000
Figure 2-31: Energy Rate Custom Units - Energy Conversion Example
Time conversion example
To find the Base per Custom relationship for a custom unit of standard cubic feet per hour
(scf/h) and a base unit of standard cubic feet per minute (StdCuft/min), input the
following:
Custom Unit = scf/h Base Unit = StdCuft/min
Because:
1 h (Hour) = 60 min (Minutes)
Then: 1 scf/h = × = 0.016667 StdCuft/min
1 scf/h = 0.016667 StdCuft/min
Therefore:
Base per Custom = = = 0.016667
Reference Manual Configuration
00809-0100-4803 October 2018
Emerson.com/Rosemount 51