8
Reference Manual
00809-0100-4021, Rev GC
Section 1: Introduction
April 2014
Introduction
1.5 Confirm HART revision capability
If using HART based control or asset management systems, confirm the HART capability of
those systems prior to transmitter installation. Not all systems are capable of communicating
with HART Revision 7 protocol. This transmitter can be configured for either HART Revision 5 or
Revision 7.
Switch HART revision mode
If the HART configuration tool is not capable of communicating with HART Revision 7, the
3144P will load a Generic Menu with limited capability. The following procedures will switch the
HART revision mode from the Generic Menu.
1. Manual Setup>Device Information>Identification>Message
a. To change to HART Revision 5, Enter: “HART5” in the Message field
b. To change to HART Revision 7, Enter: “HART7” in the Message field