Appendix B: Product Certifications
April 2014
Product Certifications
Tables
Table B-2. Process Temperatures
Table B-3. Entity Parameters
Temperature Ambient Process temperature w/o LCD display cover (°C)
class temperature
No ext. 3 in. 6 in. 9 in.
T6
– 50 °C to +40 °C
55 55 60 65
T5
– 50 °C to +60 °C
70 70 70 75
T4
– 50 °C to +60 °C
100 110 120 130
T3
– 50 °C to +60 °C
170 190 200 200
T2
– 50 °C to +60 °C
280 300 300 300
T1
– 50 °C to +60 °C
440 450 450 450
Fieldbus/Profibus HART 5
Voltage U
i
(V) 30 30
Current I
i
(mA) 300 300
Power P
i
(W) 1 1.3
Capacitance C
i
(nF) 5 2.1
Inductance L
i
(mH) 0 0