Design and function qTOWER iris
12
The voltage selector switch is located on the bottom of the device, behind a cover. It can
be used to adjust the operating voltage to the mains voltage.
Fig.4 Underside of the device with voltage selector switch
The following accessories are included in the delivery scope of the device:
¡ Power cable
¡ Connection cable for connecting to the PC
¡ CD or USB stick with qPCRsoft software, with manual for software and device
¡ Operating manual and software manual (print version)
¡ Packaging and packaging instructions
Fig.5 Mains and connection cable
Only use the mains cable supplied or a mains cable with the same specifications.
4.2 Function
4.2.1 Fluorescence spectrometer
The detector unit for a thermal block with 96 wells is a patented 8-channel epi-fluores-
cence photometer with fiber multiplexer and a mechanical scanning device.
The detector unit for a thermal block with 384 wells is a patented 16-channel epi-fluo-
rescence photometer with fiber multiplexer and a mechanical scanning device.
The device is an open platform for real-time PCR and supports both intercalating dyes as
well as individual sensors and kits from various manufacturers. The device can be used
in various applications, such as expression analyses, genotyping and the detection of
pathogens.
Light source
A long-life, sturdy five-color LED (blue, green, white, red, and UV) in the device is used
as an excitation light source for the emission of fluorescent dyes. The light source does
not require any preheating time.