Maintenance and care qTOWER iris
24
8.1.1 Cleaning the housing
If the device becomes contaminated during daily use, cleaning with a damp cloth is suffi-
cient.
Only wipe the housing with a soft, clean cloth which may be wetted with a commercially
available neutral cleaning agent if necessary.
8.1.2 Cleaning the sample block
WARNING
Risk of eye injury!
Wear safety goggles when blasting the sample block with compressed air.
Dust or reagent residue in the sample block can cause an increase in the background sig-
nal.
¡ Use compressed air to blow out the recesses in the sample block.
¡ To remove reagent residue, fill the affected wells with max. 20µL of distilled water
or ethanol. Let the liquid act for approx. 1min before sucking it out again. Repeat
the process until the background signal is in the normal range.
8.1.3 Disinfecting the device
¡ Avoid contamination by handling samples carefully.
¡ Wipe spilled samples or reagents immediately with an absorbent cloth or piece of
paper.
¡ The device cannot be decontaminated in its entirety. Take special care when using
the instrument to analyze infectious material.
¡ Remove visible contamination immediately with suitable means. Do not allow sol-
vents to enter the device.
¡ The sample block is also suitable for wipe and spray disinfection. The only suitable
cleaning method for the housing is wipe disinfection.
Device part Recommended disinfectants Provider
Housing Descosept Spezial Dr. Schuhmacher GmbH
Sample chamber Descosept Spezial
Meliseptol HBV (tissues)
Dr. Schuhmacher GmbH
B. Braun
Observe the efficacy spectrum of the listed disinfectants with regard to the customer-
specific decontamination requirements!
¡ Only disinfection agents containing ethanol or isopropanol are approved. Disinfec-
tion agents with ingredients or concentrations other than those in the recom-
mended disinfection agents may result in insufficient decontamination or damage to
the device.
¡ Use disinfection agents other than those presented here only after consultation with
Analytik Jena.
¡ If the device must be sent back to Analytik Jena for servicing, first perform decon-
tamination and document this (→ "Return" 27).
See also
2 Return [}27]