qTOWER iris Specifications
33
Electrical connection require-
ments
Operating voltage 100/115/230V(AC)±10%
Line frequency 50/60Hz
Power consumption ≤850W
Device fuses 2x10AT/250V
Minimum PC requirements
Processor Intel Core 3 or better
Working memory 2048MB RAM
Display resolution Min. 1280x1024pixels
Operating system Windows 10 or higher
Interfaces ¡ USB port
¡ Ethernet interface
Additional interfaces for models with inte-
grated tablet
¡ USB port for data export on the front of
the device
¡ Service port on the rear of the device
Software
qPCRsoft Control and analysis program
Analysis methods ¡ Absolute quantification
¡ Relative quantification
¡ ΔΔCt method
¡ Allelic discrimination
¡ Efficiency calculation
¡ DNA melting curves
¡ POS/NEG analysis in the end point
Export functions Excel, CSV, LIMS, GenEx, qBase+, GeneIO
Integrated tablet
Tablet 10", color, with touch screen
Data connection USB for data transmission via USB stick
Operating system Windows 10
11.2 Ambient conditions
Operation Transport, storage
Temperature range +15°Cto+35°C -10°Cto+55°C
Max. humidity 70% 10%to30%
Use desiccant!
Max. permissible height 2000m
Air pressure 0.7to1.06bar
Work environment Only suitable for operation in rooms
Pollution degree of the in-
tended environment
2