Cerabar M, Deltabar M, Deltapilot M Commissioning with an operating menu (onsite display/FieldCare)
Endress+Hauser 127
Expert → Measurement → Flow (Deltabar M)
Parameter name Description
Flow type (044)
Options
Select the flow type.
Options:
• Volume process cond. (volume under operating conditions)
• Volume norm. cond. (norm volume under norm conditions in Europe:
1013.25 mbar and 273.15 K (0°C))
• Volume std. cond. (standard volume under standard conditions in the USA:
1013.25 mbar (14.7 psi) and 288.15 K (15°C/59°F))
• Mass (mass under operating conditions)
• Flow in %
Factory setting:
Volume process conditions
Mass flow unit (045)
Options
Select mass flow unit.
When a new flow unit is selected, all flow-specific parameters are converted and
displayed with the new unit within a flow type. When the flow type is changed,
conversion is not possible.
Prerequisite:
•"Flow type (044)" = Mass
Options:
• g/s, kg/s, kg/min, kg/h
• t/s, t/min, t/h, t/d
•oz/s, oz/min
• lb/s, lb/min, lb/h
• ton/s, ton/min, ton/h, ton/day
Factory setting:
kg/s
Norm. flow unit (046)
Options
Select norm flow unit.
When a new flow unit is selected, all flow-specific parameters are converted and
displayed with the new unit within a flow type. When the flow type is changed,
conversion is not possible.
Prerequisite:
•"Flow type (044)" = Volume norm. cond.
Options:
•Nm
3
/s, Nm
3
/min, Nm
3
/h, Nm
3
/d
Factory setting:
Nm
3
/s
Std. flow unit (047)
Options
Select standard flow unit.
When a new flow unit is selected, all flow-specific parameters are converted and
displayed with the new unit within a flow type. When the flow type is changed,
conversion is not possible.
Prerequisite:
•"Flow type (044)" = Volume std. conditions
Options:
•Sm
3
/s, Sm
3
/min, Sm
3
/h, Sm
3
/day
• SCFS, SCFM, SCFH, SCFD
Factory setting:
Sm