Cerabar M, Deltabar M, Deltapilot M Commissioning with an operating menu (onsite display/FieldCare)
Endress+Hauser 129
Expert Measurement Sensor trim
8.11.3 Communication
Expert Communication FF Info
Expert Communication Resource Block (only with FieldCare)
See ä 166 ff.
Expert Communication Transducer Blocks (only with FieldCare)
See ä 176 ff.
Parameter name Description
Lo trim measured (129)
Display
Displays the reference pressure present to be accepted for the lower calibration
point.
Hi trim measured (130)
Display
Displays the reference pressure present to be accepted for the upper calibration
point.
Lo trim sensor (131)
Display
Internal service parameter.
Hi trim sensor (132)
Display
Internal service parameter.
Parameter name Description
Device type code (236)
Display
The "Device type code (236)" is the unique device ID in the control system or the FF
bus.
It consists of the manufacturer ID (452B48), device type number and device serial
number.
Example:
Deltabar M: 452B481021-6B032A0109D
Device revision (237)
Display
Displays the revision or version of a complete device (HW+SW).
Example:
1
Device address (244)
Display
Displays the device address currently configured and valid.
Factory setting:
247
Device class (245)
Display
Displays the device class currently configured.
The device can be configured as a "Basic device" or "Link master".
Factory setting:
Basic device