Deltabar S
26 Endress+Hauser
Pressure specifications The maximum pressure for the measuring device is dependent on the lowest-rated element with regard to
pressure, see the following sections for this:
– → page 12, section "Measuring range"
– → page 23, section "Oxygen applications"
– → chapter "Mechanical construction".
Observe pressure-temperature dependency.
Note!
• The MWP (maximum working pressure) is specified on the nameplate. This value refers to a reference
temperature of 68°F (20°C) or 100°F for ANSI flanges.
• Test pressure (Over pressure limit OPL) = MWP (nameplate) x 1.5.
• The pressure values permitted at higher temperatures can be found in the following standards:
– EN 1092-1: 2001 Tab. 18
– ASME B 16.5a – 1998 Tab. 2-2.2 F316
– ASME B 16.5a – 1998 Tab. 2.3.8 N10276
– JIS B2201
Mechanical construction
Housing dimensions, optional
display on the side (T14)
P01-xMD7xxxx-06-00-xx-xx-000
Front view, left-hand side view, top view
Housing dimensions, optional
display on the top (T15)
P01-xMD7xxxx-06-00-xx-xx-001
Front view, left-hand side view, top view