EasyManua.ls Logo

Endress+Hauser Deltabar S PMD75 - Page 160

Endress+Hauser Deltabar S PMD75
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter description (Local operation and FieldCare) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
160 Endress+Hauser
CUST. UNIT FACT. V (607)
Entry
Enter the conversion factor for a customer-specific volume unit.
The conversion factor must be entered in relation to the SI unit "m
3
".
See also CUSTOMER UNIT V.
Prerequisite:
LIN. MEASURAND = Volume, UNIT VOLUME = User unit
Example:
You want the measured value to be displayed in "buckets".
MEASURED VALUE = 0.01 m3 i 1 bucket
Entry CUSTOMER UNIT V: bucket
Entry CUST. UNIT FACT. V: 100
Result: MEASURED VALUE = 1 bucket
Factory setting:
1.0
MASS UNIT (709)
Selection
Select the mass unit.
Prerequisite:
LIN. MEASURAND = Mass
Options:
•g
•kg
•t
•oz
•lb
•ton
•User unit, see also the following parameter descriptions for CUSTOMER UNIT M
and CUST. UNIT. FACT. M
!
Note!
Following a change in the unit, the digital output value of the Analog Input value OUT
and the local operation or the MEASURED VALUE no longer display the same value.
See also parameter descriptions for XD SCALE, OUT SCALE and OUT value.
Factory setting:
kg
CUSTOMER UNIT M (704)
Entry
Enter text (unit) for customer-specific mass unit.
You can enter a maximum of eight alphanumeric characters here.
See also CUST. UNIT. FACT. M.
Prerequisite:
LIN. MEASURAND = Mass, MASS UNIT = User unit
!
Note!
Only the first five characters are shown on the local operation. For example, if the
customer-specific unit "crates" was selected, only "crate" would be displayed.
If the unit contains a slash, up to eight characters can be shown on the local operation.
The maximum number of characters in the counter is again limited to five. For example if
the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed.
FieldCare accepts units with eight characters at most. The slash counts as one character.
Factory setting:
_ _ _ _ _ _ _ _
CUST. UNIT FACT. M (703)
Entry
Enter the conversion factor for a customer-specific mass unit.
The conversion factor must be entered in relation to the SI unit "kg".
See also CUSTOMER UNIT M.
Prerequisite:
LIN. MEASURAND = Mass, MASS UNIT = User unit
Example:
You want the measured value to be displayed in "buckets".
MEASURED VALUE = 10 kg i 1 bucket
Entry CUSTOMER UNIT M: bucket
Entry CUST. UNIT FACT. M: 0.1
Result: MEASURED VALUE = 1 bucket
Factory setting:
1.0
Table 11: OPERATING MENU SETTINGS BASIC SETUP "Level", LEVEL MODE "Linear"