Parameter description (Local operation and FieldCare) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
186 Endress+Hauser
P01-xxxxxxxx-19-xx-xx-en-091
Fig. 52: LINEARIZATION function group for local operation
1
1
EDITOR TABLE
EDITOR TABLE
MEASURINGTABLE
Editor Table
Measuring Table
EDITOR TABLE
TANK CONTENT MIN
LIN. EDIT MODE
TANK DESCRIBTION
TANK CONTENT MAX
TABLE SELECTION
MEASURINGTABLE
LANGUAGE
DISPLAYSETTINGS
POSITION ADJUSTMENT
OPERATING MENU
GROUP SELECTION
Measured value
LINEARIZATION
Level measuring
mode only
1) Displayvia on-site display only
Table 18: GROUP SELECTION OPERATING MENU SETTINGS LINEARIZATION
– Local operation
Parameter name Description
Prerequisite:
MEASURING MODE = Level ( ä 134).
LEVEL MODE = Pressure linearized or height linearized ( ä 156).
Note:
See also ä 105 ff, Section 9 "Level measurement (via local operation and FieldCare)".
TANK CONTENT MIN (759)
Entry
Enter the minimum tank contents to be expected.
The input limits for the calibration (editing limits) are derived from the value entered. The
closer the value entered corresponds to the minimum tank content to be expected, the
more accurate the measurement result.
Factory setting:
0.0
TANK CONTENT MAX
(713)
Entry
Enter the maximum tank contents to be expected.
The input limits for the subsequent calibration (editing limits) are derived from the value
entered. The closer the value entered corresponds to the maximum tank content to be
expected, the more accurate the measurement result.
Factory setting:
100.0
TABLE SELECTION (808)
Selection
Select table.
The device works with a measuring and an editor table. The measuring table is used to
calculate the measured value. To make sure measuring also runs properly when entering
a new table, there is another table, the editor table, for entering new values.
Options:
•View meas. table
•Editor table
Factory setting:
View meas. table