Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Troubleshooting
Endress+Hauser 211
116 (W116) Warning
C
C>Download error,
repeat download
17 0 The file is defect. Use another file. 37
During the download, the data
are not correctly transmitted to
the processor, e.g. because of
open cable connections, spikes
(ripple) on the supply voltage or
electromagnetic effects.
Check cable connection PC –
transmitter.
Block off electromagnetic effects
or eliminate sources of
disturbance.
Perform reset (code 7864) and
recalibrate the device.
–Repeat download.
120 (E120) Error
C
C>Sensor low pressure 17 0 Pressure too low. Increase pressure until message
disappears.
31
Sensor defect. Replace sensor.
121 (A121) Alarm
B
B>Checksum error in
factory segment of
EEPROM
23 11 Main electronics defect. Replace main electronics. 5
122 (A122) Alarm
B
B>Sensor not connected 20 5 Cable connection sensor –main
electronics disconnected.
Check cable connection and
repair if necessary.
14
Electromagnetic effects are
greater than specifications in the
technical data.
See Technical Information
TI00382P (Deltabar S),
TI00383P (Cerabar S) or
TI00416P (Deltapilot S).
Block off electromagnetic effects
or eliminate source of
disturbance.
Main electronics defect. Replace main electronics.
Sensor defect. Replace sensor.
130 (A130) Alarm
B
B>EEPROM is defect. 23 11 Main electronics defect. Replace main electronics. 10
131 (A131) Alarm
B
B>Checksum error in
EEPROM: min/max
segment
23 11 Main electronics defect. Replace main electronics. 9
132 (A132) Alarm
B
B>Checksum error in
totalizer EEPROM
23 11 Main electronics defect. Replace main electronics. 7
133 (A133) Alarm
B
B>Checksum error in
History EEPROM
23 11 An error occurred when
writing.
Perform reset (code 7864) and
recalibrate the device.
8
Main electronics defect. Replace main electronics.
134 (W134) Warning
C
C>EEPROM lifetime
WARNING
17 0 Writing too often to EEPROM. Reduce write accessing to
EEPROM.
55
135 (A135) Alarm
B
B>Checksum error in
EEPROM FF segment
23 11 Main electronics defect. Replace main electronics. 10
602 (W602) Warning
C
C>Linearization curve
not monoton
19 7 The linearization table is not
monotonic increasing.
Add to or correct linearization
table. Then accept linearization
table again.
57
604 (W604) Warning
C
C>linearization table
invalid. Min. 2 points
19 7 The linearization table consists
of less than 2 points.
Add to linearization table.
Accept linearization table again
if necessary.
58
613 (W613) Warning
I
I>Simulation is active 17 3 Simulation is switched on, i.e.
the device is not measuring at
present.
Switch off simulation. 60
Code Error
type/
NA 64
Message/description XD_
ERROR
Value
BLOCK_
ERROR
Bit
Cause Measure Prio-
rity

Table of Contents