Troubleshooting Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
212 Endress+Hauser
700 (W700) Warning
C
C>Last configuration not
stored
23 11 – An error occurred when writing
or reading configuration data or
the power supply was
disconnected.
– Perform reset (code 7864) and
recalibrate the device.
53
– Main electronics defect. – Replace main electronics.
702 (W702) Warning
C
C>HistoROM data not
consistent.
17 11 – Data were not written correctly
to the HistoROM, e.g. if the
HistoROM was detached during
the writing process.
–Repeat upload.
– Perform reset (code 7864) and
recalibrate the device.
54
– HistoROM does not have any
data.
– Copy suitable data to the
HistoROM. (→ See also
Operating Instructions
BA00301P (Deltabar S) or
BA00302P (Cerabar S) or
BA00372P (Deltapilot S),
"Copying configuration data"
section.)
703 (A703) Alarm
B
B>Measurement error 20 5 – Fault in the main electronics. – Briefly disconnect device from
the power supply.
23
– Main electronics defect. – Replace main electronics.
704 (A704) Alarm
B
B>Measurement error 20 5 – Fault in the main electronics. – Briefly disconnect device from
the power supply.
13
– Main electronics defect. – Replace main electronics.
705 (A705) Alarm
B
B>Measurement error 20 5 – Fault in the main electronics. – Briefly disconnect device from
the power supply.
22
– Main electronics defect. – Replace main electronics.
706 (W706) Warning
C
C>Configuration in
HistoROM and device
not identical
23 11 – Configuration (parameters) in
the HistoROM and in the device
not identical.
– Copy data from the device to
the HistoROM. (→ See also
Operating Instructions
BA00301P (Deltabar S) or
BA00302P (Cerabar S) or
BA00372P (Deltapilot S),
"Copying configuration data"
section.)
– Copy data from the HistoROM
to the device. (→ See also
Operating Instructions
BA00301P (Deltabar S) or
BA00302P (Cerabar S) or
BA00372P (Deltapilot S),
"Copying configuration data"
section.) The message remains if
the HistoROM and the device
have different software versions.
The message disappears if you
copy the data from the device to
the HistoROM.
– Device reset codes such as 7864
(onsite) or 40864 (bus) do not
have any effect on the
HistoROM.
That means that if you perform a
reset, the configurations in the
HistoROM and in the device
may not be the same.
59
Code Error
type/
NA 64
Message/description XD_
ERROR
Value
BLOCK_
ERROR
Bit
Cause Measure Prio-
rity