Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Troubleshooting
Endress+Hauser 213
707 (A707) Alarm
B
B>X-VALUE
(LINEARIZATION_TAB
LE_X_VALUE) of lin.
table out of edit limits.
18 7 At least one X-VALUE
(LINEARIZATION_TABLE_X_
VALUE) in the linearization
table is either below the value
for Scale In/SCALE_IN, EU at
0%/EU_0, HYDR. PRESS MIN.
or Level Min/LEVEL_MIN or
above the value for Scale In/
SCALE_IN, EU at 100%/
EU_100, HYDR. PRESS. MAX.
or Level Max/LEVEL_MAX.
Carry out calibration again.
(See also these Operating
Instructions, Section 4 and
Section 8)
38
710 (W710) Warning
C
B>Set span too small.
Not allowed.
18 7 Values for calibration (e.g.
lower-range value and upper-
range value) are too close
together.
Adjust calibration to suit sensor.
( ä 54 and ä 198,
parameter description for
Calibration Minimum Span/
CAL_MIN_SPAN.)
50
The sensor was replaced and the
customer-specific configuration
does not suit the sensor.
Adjust calibration to suit sensor.
Replace sensor with a suitable
sensor.
Unsuitable download carried
out.
Check configuration and
perform download again.
713 (A713) Alarm
B
B>100% Point
(HUNDRED_PERCENT_
VALUE) level out of edit
limits
18 7 The sensor was replaced. Carry out calibration again. 39
715 (E715) Error
C
C>Sensor over
temperature
17 0 The temperature measured in
the sensor is greater than the
upper nominal temperature of
the sensor. ( ä 77 and
ä 199, parameter
description for Tmax. Sensor/
SENSOR_TEMP_HIGH_
LIMIT)
Reduce process temperature/
ambient temperature.
33
Unsuitable download carried
out.
Check configuration and
perform download again.
716 (A716) Alarm
B
B>Process isolating
diaphragm broken
20 5 Sensor defect. Replace sensor. 25
717 (E717) Error
C
C>Transmitter over
temperature
17 0 The temperature measured in
the electronics is greater than
the upper nominal temperature
of the electronics (+88 °C (190
°F)).
Reduce ambient temperature. 35
Unsuitable download carried
out.
Check configuration and
perform download again.
718 (E718) Error
C
C>Transmitter under
temperature
17 0 The temperature measured in
the electronics is smaller than
the lower nominal temperature
of the electronics (–43 °C (-45
°F)).
Increase ambient temperature.
Insulate device if necessary.
36
Unsuitable download carried
out.
Check configuration and
perform download again.
Code Error
type/
NA 64
Message/description XD_
ERROR
Value
BLOCK_
ERROR
Bit